Norovirus (GⅡ) RT-PCR Detection Kit

Short Description:

It is suitable for the detection of Norovirus (GⅡ) in shellfish, raw vegetables and fruits, water, feces, vomit and other specimens.


Product Detail

Product Tags

Product Name

Norovirus (GⅡ) RT-PCR Detection Kit (Lyophilized)

Size

48tests/kit, 50tests/kit

Intended Use

It is suitable for the detection of Norovirus (GⅡ) in shellfish, raw vegetables and fruits, water, feces, vomit and other specimens. Nucleic acid extraction should be carried out by nucleic acid extraction kit or direct pyrolysis method according to different specimen types.The kit is an ALL-READY PCR SYSTEM(Lyophilized), which contains the DNA amplification enzyme, reverse transcriptase,reaction buffer, specific primers and probes required for fluorescent PCR detection.

Product Contents

Components Package specification Ingredient
Norovirus (GⅡ) PCR Mix 1 ×bottle                  (Lyophilized powder)  50Test dNTPs, MgCl2, Primers, Probes,Reverse Transcriptase,Taq DNA polymerase
6×0.2ml 8 well-strip tube(Lyophilized) 48Test
Positive Control 1*0.2ml tube (lyophilized)  10Tests

Plasmid containing Norovirus (GⅡ) specific fragments

Dissolving solution 1.5 ml Cryotube 500uL /
Negative Control 1.5 ml Cryotube 200uL 0.9%NaCl

Storage & Shelf Life

(1)The kit can be transported at room temperature.

(2)The shelf life is 18 months at -20℃ and 12 months at 2℃~30℃.

(3)See the label on kit for the production date and expiration date.

(4)The lyophilized powder version reagent should be stored at -20℃ after dissolution and the repeated freeze -thaw should be less than 4 times.

Instruments

GENECHECKER UF-150, UF-300 real-time fluorescence PCR instrument.

Operation Diagram

a) Bottle version:

1

b)  8 well-strip tube version:

2

Pcr Amplification

Recommended Setting

Step     Cycle Temperature (℃)   Time Fluorescence channel
1 1 50 8min  
2 1 95 2min  
3 40 95 5s  
60 10s Collect FAM fluorescence

*Note:  The signals of FAM fluorescence channels will be collected at 60℃.

Interpreting Test Results

Channel

Interpretation of results

FAM Channel

Ct≤35

Norovirus (GⅡ) Positive

Undet

Norovirus (GⅡ) Negative

35<Ct≤40

Suspicious resut, retest*

*If the retest result of FAM channel has a Ct value ≤40 and shows typical “S” shape amplification curve, the result is interpreted as positive, otherwise it is negative.